Tag Archives: spectrophotometry

Ultraviolet spectrophotometric methods in dissolution test

J. Pharm. Pharmacogn. Res., vol. 10, no. 5, pp. 847-856, September-October 2022.

Original Article

Development and validation of ultraviolet spectrophotometric methods for lamivudine and zidovudine quantification in dissolution test

[Desarrollo y validación de métodos espectrofotométricos ultravioleta para la cuantificación de lamivudina y zidovudina en pruebas de disolución]

Enma Perez-Chauca1*, Pedro M. Alva-Plasencia2, Humberto Gomes Ferraz3

1Unidad de Posgrado en Farmacia y Bioquímica, Facultad de Farmacia y Bioquímica, Universidad Nacional de Trujillo, Trujillo, 13011, Perú.

2Departamento de Farmacotecnia, Facultad de Farmacia y Bioquímica, Universidad Nacional de Trujillo, Trujillo, 13011, Perú.

3Faculdade de Ciências Farmacêuticas, Universidad de São Paulo, Av. Prof. Lineu Prestes 580-Cidade universitária, 05508-900, São Paulo, Brasil.  

*E-mail: eperez@unitru.edu.pe

Abstract

Context: In the development of a good analytical method, the selection of appropriate conditions for quantifying drugs is essential. The method validation should be determined before conducting the dissolution test.

Aims: To develop and validate two analytical methods by UV spectrophotometry to quantify lamivudine and zidovudine in dissolution test.

Methods: The dissolution conditions were 900 mL of dissolution medium (pH 1.2, pH 4.5, and pH 6.8), using paddles, at 75 rpm, and sampling time points of 5, 10, 15, 20, and 30 minutes. The analytical methods were developed by scanning analytical solutions in the UV region. The proposed methods were validated in accordance with the ICH guideline and the United States Pharmacopoeia.

Results: Lamivudine showed wavelengths at 280 nm, 273 nm, and 270 nm in buffers at pH 1.2, pH 4.5, and pH 6.8, respectively; while zidovudine showed a wavelength at 266 nm in all three buffers. Regression analysis confirmed linearity (r2>0.998). Placebos and diluents showed no analytical interference. LOD and LOQ were lower than the linearity range. Recovery percentages were within 95% to 105%. RSD values were below 2% and 7% for repeatability and intermediate precision, respectively. No changes introduced modified the method response. The analytical solutions were stable until 24 hours. The results showed that Whatman No 41, No 42, and PVDF 0.45 µm filters can be used.

Conclusions: The proposed analytical methods demonstrated linearity, specificity, accuracy, and precision. In addition, LOD and LOQ, robustness, stability of the analytical solution, and filter test showed satisfactory results.

Keywords: analytical methods; drug dissolution; reverse transcriptase inhibitors; spectrophotometry.

jppres_pdf_free

Resumen

Contexto: En el desarrollo de un buen método analítico, la selección de las condiciones adecuadas para cuantificar los fármacos es esencial. La validación del método debe determinarse antes de realizar las pruebas de disolución.

Objetivos: Desarrollar y validar dos métodos analíticos por espectrofotometría UV para cuantificar lamivudina y zidovudina en pruebas de disolución.

Métodos: Las condiciones de disolución fueron de 900 mL de medio de disolución (pH 1,2; pH 4,5 y pH 6,8), utilizando paletas, a 75 rpm, y puntos de tiempo de muestreo de 5, 10, 15, 20 y 30 minutos. Los métodos analíticos se desarrollaron mediante el escaneo de las soluciones analíticas en la región UV. Los métodos propuestos fueron validados de acuerdo con la directriz ICH y la Farmacopea de los Estados Unidos.

Resultados: Lamivudina mostró longitudes de onda a 280 nm, 273 nm y 270 nm en tampones a pH 1,2; pH 4,5 y pH 6,8, respectivamente; mientras que zidovudina mostró una longitud de onda de 266 nm en los tres tampones. El análisis de regresión confirmó la linealidad (r2>0,998). Los placebos y diluyentes no mostraron interferencia analítica. El LOD y LOQ fueron más bajos que el rango de linealidad. Los porcentajes de recuperación estuvieron dentro del 95% al 105%. Los valores de RSD fueron inferiores al 2% y 7% para la repetibilidad y la precisión intermedia, respectivamente. Ningún cambio introducido modificó la respuesta del método. Las soluciones analíticas fueron estables hasta las 24 horas. Los resultados mostraron que se pueden utilizar filtros Whatman No 41, No 42 y PVDF 0,45 μm.

Conclusiones: Los métodos analíticos propuestos demostraron linealidad, especificidad, exactitud y precisión. Además, LD y LC, la robustez, la estabilidad de la solución analítica y la prueba de filtro mostraron resultados satisfactorios.

Palabras Clave: métodos analíticos; disolución de fármacos; inhibidores de la transcriptasa inversa; espectrofotometría.

jppres_pdf_free

Citation Format: Perez-Chauca E, Alva-Plasencia PM, Ferraz HG (2022) Development and validation of ultraviolet spectrophotometric methods for lamivudine and zidovudine quantification in dissolution test. J Pharm Pharmacogn Res 10(5): 847–856.
References

AEFI (2001) Validación de métodos analíticos. Barcelona: Asociación Española de Farmacéuticos de la Industria.

BP (2018a) Lamivudine Tablets. In: The British Pharmacopoeia. The stationary office on behalf of the Medicines and Healthcare Products Regulatory Agencies, London.

BP (2018b) Zidovudine Tablets. In: The British Pharmacopoeia. The stationary office on behalf of the Medicines and Healthcare Products Regulatory Agencies, London.

Chen Y, Gao Z, Duan JZ (2017) Dissolution Testing of Solid Products. In: Developing Solid Oral Dosage Forms. Elsevier, pp 355–380.

Fernandes C, Moreira de Campos LM, Pianetti GA (2003) Validação de metodologia para doseamento e estudo de equivalência farmacêutica de comprimidos de lamivudina 150 mg. Braz J Pharm Sci (4): 381–389.

Friedel HD, Brown CK, Barker AR, Buhse LF, Keitel S, Kraemer J, Morris JM, Reppas C, Sperry DC, Sakai-Kato K, Stickelmeyer MP, Shah VP (2018) FIP guidelines for dissolution testing of solid oral products. J Pharm Sci 107(12): 2995–3002.

Gao Q, Sanvordeker DR (2014) Analytical Methods Development and Methods Validation for Oral Solid Dosage Forms. In: Generic Drug Product Development. CRC Press, pp 31-50.

Gray VA, Rosanske TW (2014) Dissolution. In: Specification of Drug Substances and Products: Development and Validation of Analytical Methods. Elsevier, pp 245-263.

Hwisa NT, Adiki SK, Katakam P, Chandu BR (2013) Design of dissolution media for in-vitro bioequivalence testing of Lamivudine. J App Pharm Sci 3(6): 106–110.

ICH (2005) Q2 (R1) Validation of analytical procedures: text and methodology. ICH harmonized tripartite guidelines. International Conference on Harmonisation of Technical Requirements for Registration of Pharmaceuticals for Human Use.

Mandloi DK, Tyagi PK, Rai VK, Dey S, Ashada RK, Mohanraj P (2009) Method development and validation of RP- HPLC in the application of in-vitro dissolution study of Lamivudine in bulk drug and tablet formulation. J Chem Pharm Res 1(1): 286–296.

Ozturk N, Kaynak MS, Sahin S (2015) Comparison of dissolution profiles of commercially available lamivudine tablets. Dissolution Technol 22(4): 38–43.

Reddy NHS, Patnala S, Löbenberg R, Kanfer I (2014) In vitro dissolution of generic immediate-release solid oral dosage forms containing BCS class I drugs: Comparative assessment of metronidazole, zidovudine, and amoxicillin versus relevant comparator pharmaceutical products in South Africa and India. AAPS PharmSciTech 15(5): 1076-1086.

Reppas C, Friedel HD, Barker AR, Buhse LF, Cecil TL, Keitel S, Kraemer J, Morris JM, Shah VP, Stickelmeyer MP, Yomota C, Brown CK (2014) Biorelevant in vitro performance testing of orally administered dosage forms – Workshop report. Pharm Res 31(7): 1867–1876.

Soares K, Rediguieri C, Souza J, Serra C, Abrahamsson B, Groot D, Kopp S, Langguth P, Polli J, Shah V, Dressman J (2013) Biowaiver Monographs for immediate-release solid oral dosage forms: Zidovudine (Azidothymidine). J Pharm Sci 102(8): 2409–2423.

Strauch S, Jantratid E, Dressman J, Junginger H, Kopp S, Midha K, Shah V, Stavchansky S, Barends D (2011) Biowaiver monographs for immediate release solid oral dosage forms: Lamivudine. J Pharm Sci 100(6): 2054–2063.

Stuart AV, Zuo J, Löbenberg R (2014) Investigating the dissolution profiles of amoxicillin, metronidazole, and zidovudine formulations used in Trinidad and Tobago, West Indies. AAPS PharmSciTech 15(5): 1060–1069.

USP (2019a) Lamivudina, tabletas. In: Pharmacopoeia of the United States of America, 42nd edn. Rockville: The United States Pharmacopeial Convention, pp 2593–2595.

USP (2019b) Zidovudina, tablets. In: Pharmacopoeia of the United States of America, 42nd edn. Rockville: The United States Pharmacopeial Convention, pp 4708-4709.

USP (2019c) <1092> Procedimiento de Disolución: Desarrollo y Validación. In: Pharmacopoeia of the United States of America, 42nd edn. Rockville: The United States Pharmacopeial Convention, pp 7685–7707.

USP (2019d) <1225> Validación De Procedimientos Farmacopeicos. In: Pharmacopoeia of the United States of America, 42nd edn. Rockville: The United States Pharmacopeial Convention, pp 8387-8393.

Wang Q, Ma D, Higgins JP (2006) Analytical method selection for drug product dissolution testing. Dissolution Technol 13(3): 6-13.

Wang Y, De Clercq E, Li G (2019) Current and emerging non-nucleoside reverse transcriptase inhibitors (NNRTIs) for HIV-1 treatment. Expert Opin on Drug Metab and Toxicol 15(10): 813–829.

WHO (2017) Multisource (generic) pharmaceutical products: guidelines on registration requirements to establish interchangeability. Annex 6. In: WHO Expert Committee on Specifications for Pharmaceutical Preparations: fifty-first report.

© 2022 Journal of Pharmacy & Pharmacognosy Research (JPPRes)

UV spectrophotometric validation of losartan tablets



J Pharm Pharmacogn Res 10(2): 310-317, 2022.

Original Article

Validation of a UV spectrophotometric method to quantify losartan potassium in tablets from the dissolution test at pH 1.2, 4.5 and 6.8

[Validación de un método espectrofotométrico UV para cuantificar losartán potásico en tabletas a partir de la prueba de disolución a pH 1,2; 4,5 y 6,8]

Antonella S. Araujo-Fernandez1, José C. Uribe-Villarreal2, Enma Perez-Chauca2, Pedro M. Alva-Plasencia1, Olga E. Caballero-Aquiño3, Mayar L. Ganoza-Yupanqui3*

1Departamento de Farmacotecnia, Facultad de Farmacia y Bioquímica, Universidad Nacional de Trujillo, Trujillo, 13011, Perú.

2Unidad de posgrado en Farmacia y Bioquímica, Facultad de Farmacia y Bioquímica, Universidad Nacional de Trujillo, Trujillo, 13011, Perú.

3Departamento de Farmacología, Facultad de Farmacia y Bioquímica, Universidad Nacional de Trujillo, Trujillo, 13011, Perú.

*E-mail: mganoza@unitru.edu.pe

Abstract

Context: The validation of a method is synonymous with the quality of the obtained results. The dissolution test is an analytical technique to evaluate the quality and stability of drugs during their development.

Aims: To evaluate whether the UV spectrophotometric method for the quantification of losartan potassium in tablets from the dissolution test at pH 1.2, 4.5 and 6.8 meet with the validation parameters.

Methods: The determination and evaluation of validation parameters were carried out under the guidelines of the regulatory entities. Linearity and range, accuracy, precision, specificity, limits of detection and quantification, robustness, stability of the sample solution were evaluated, and the filter test was added. All data obtained were subject to an analysis of variance and t-student analysis with a confidence level of 95% (α = 0.05).

Results: The UV spectrophotometric method meets the acceptance criteria for each validation parameter. Likewise, it was identified that the prepared solutions were stable at pH 6.8 for 24 hours; however, they were not stable at pH 1.2 and 4.5.

Conclusions: The method meets with the validation criteria and is suitable to be used for quantifying samples obtained from the losartan potassium tablet dissolution test.

Keywords: validation; losartan; spectrophotometry; dissolution test.

Resumen

Contexto: La validación de un método es sinónimo de calidad en los resultados obtenidos. La prueba de disolución es una técnica analítica para evaluar la calidad y estabilidad de los medicamentos durante su desarrollo.

Objetivos: Evaluar si el método espectrofotométrico UV para la cuantificación de losartán potásico en tabletas a partir de la prueba de disolución a pH 1,2; 4,5 y 6,8 cumple con los parámetros de validación.

Métodos: La determinación y evaluación de los parámetros de validación fueron realizados bajo las directrices de las entidades regulatorias. Se evaluaron linealidad y rango, exactitud, precisión, especificidad, límites de detección y cuantificación, robustez, estabilidad de la solución muestra y se adicionó la prueba de filtros. Todos los datos obtenidos fueron sujetos a un análisis de varianza y análisis t-student con un nivel de confianza del 95% (α = 0,05).

Resultados: El método espectrofotométrico UV cumple con los criterios de aceptación planteados para cada parámetro de validación. Así mismo, se identificaron que las soluciones preparadas fueron estables a pH 6,8 por 24 horas; sin embargo, no fueron estables a pH 1,2 y 4,5.

Conclusiones: El método cumple con los criterios de validación y es apto para ser usado en la cuantificación de muestras obtenidas de la prueba de disolución en tabletas de losartán potásico.

Palabras Clave: validación; losartán; espectrofotometría; prueba de disolución.

This image has an empty alt attribute; its file name is jppres_pdf_free.png
Citation Format: Araujo-Fernandez AS, Uribe-Villarreal JC, Perez-Chauca E, Alva-Plasencia PM, Caballero-Aquiño OE, Ganoza-Yupanqui ML (2022) Validation of a UV spectrophotometric method to quantify losartan potassium in tablets from the dissolution test at pH 1.2, 4.5 and 6.8. J Pharm Pharmacogn Res 10(2): 310–317.

© 2022 Journal of Pharmacy & Pharmacognosy Research (JPPRes)